Measurement » iSAR » iSAR Head

 (head) Immediate SAR Scanner for SAM Evaluation

Download Product Flyer (PDF, 1.68 MB)

Item Typical (max)
Features iSAR² (head) closely simulates the area scan tests for devices operated at the ear as required by IEC62209-1, IEEE 1528, OET65, etc. The shape and thickness of the shell correspond to the ear-mouth line of the SAM head. The filling material consists of solid absorbing gel simulating head or body tissue specifications in the frequency range of 300 MHz to 6 GHz. A sensor array of 256 dipoles covers an active measurement area of 240 mm x 120 mm that is parallel to the surface and embedded 4mm into the absorbing material. The interface is via Ethernet. Various software packages are available for production line and prototype testing applications. Customized software is also offered.
Total absorbing material > 210mm x 300mm x 80mm
(> 600mm x 600mm x 80mm)
Dielectric Parameters
(deviation from IEC 62209)
< 10%
Active Measurement Area 120mm x 240mm
Number of Active Dipoles 256
Measurement Grid 15mm
Virtual Grid w/ Combiner < 3mm
Evaluation Time < 3 s
Frequency Range
 
0.4 – 3 GHz
(<0.4 – 6 GHz)
Sensitivity < 0.05 W/kg
Dynamic Range CW: 100 W/kg
Repeatability < 0.1 dB (with combiner)
Deviation from DASY5 < 1 dB
Measurement Cycles > 3 mio
Conformity EN 292


iSAR² (head) top  view


iSAR² (head) side  view


iSAR² (head) front view

         




 
© 2009 by SPEAG, Schmid & Partner Engineering AG, CH-8004 Zurich, Switzerland | SitemapContact | Login