HAC RF Extension for DASY5 AIR, PRO and NEO
Hearing Aid Compatibility (HAC) RF Extension: RF Emission Test of Wireless Devices (E- and H-field) according to ANSI-C63.19 section 4
Available for DASY5 AIR, PRO, NEO and DASY4
| Contents | Contains the necessary hardware to test wireless devices for Hearing Aid Compatibility. Results are displayed in M-class levels as required for ANSI-C63.19. | |
| Test Arch Phantom |
Enables easy and well defined positioning of the phone and validation dipoles as well as simple teaching of the robot. Dimensions: 370 x 370 x 370 mm |
|
| Phone Positioner | Supports accurate and reliable positioning of any phone. Effect on near field <+/- 0.5 dB |
|
| RF Emission Calibration Dipoles |
CD700V3 - CD2450V3 The dipoles can be used to test hearing aids as described in ANSI-C63.19. |
|
| Calibration | ISO/IEC 17025 calibration service available. | |
| Frequency Band | 800-960 MHz; 1710-2000 MHz; 2250-2650 MHz (in free space) | |
| Return Loss | > 15-18 dB over frequency band > 20 dB at calibration frequency |
|
| Options | Dipoles for other frequencies or solutions and other calibration conditions upon request | |
| Software |
Easy teaching of predefined ANSI-C63.19 measurement area
Predefined measurement procedures for compliance and validation Evaluation incorporates automatic exclusion of high-level areas Documentation ready for inclusion into compliance report |
|
| Probes | Isotropic free field H3DV6 H Probe and and ER3DV6 E Probe needed for RF field measurements. ISO/IEC 17025 calibration service available. | |
| Remarks | Base station simulator not included | |